ECM1/2B compressor condenser
Working principle:
Product description: The compressor condenser is used in the sample gas processing device of the analytical instrument to remove moisture from the sample gas. The accurate measurement of gas composition requires that the sample gas can maintain a stable dew point even in harsh environments, and the compressor condenser is used to remove the influence of moisture on the analyzer, thereby improving the measurement accuracy and extending the service life of the instrument.
Application scope: flue gas online monitoring, desulfurization and denitrification flue gas pretreatment,
Technical features:
1. Simple installation, it adopts wall-mounted installation, can be directly fixed in the cabinet, and has beautiful appearance;
2. The principle is reliable, the compressor is used for refrigeration and cooling, and the compressor runs continuously to avoid the impact of frequent starting on the life of the compressor;
3. Compact dual-path heat exchanger with special technology, dual-path system with hot and cold separation
4. High-precision control keeps the heat exchanger temperature constant at the set temperature with a control accuracy of ±0.2℃;
5. Fault alarm output: when the heat exchanger temperature is higher than the set value, an alarm signal is output
6. Environmentally friendly and pollution-free, the compressor refrigerant uses R134a, which has no pollution to atmospheric ozone, is green and environmentally friendly, and has good refrigeration effect;
Technical parameters:
1. Ambient temperature: 0~40℃
2. Temperature control accuracy: <±0.2℃
3. Intake air temperature: <80℃
4. Sample gas flow: <200L/h
5. Precooling time: 15 minutes
6. Dew point temperature: 4°C (factory setting)
7. Power supply: 220VAC 50/60Hz
8. Power: <350VA, fuse: 10A
9. Installation method: wall mounted